506.True regarding digital radiography
a. Detective quantum efficiency (DOE)
reflects the efficiency of phosphor detection and noise added to the detector
signal
b.DQE may be 100% in DR system
c.DQE of CR system and film-screen
system is close to 30%
d.Disadvantage of DR in comparison to
CR are cost and versatility
e.a DR system have the same flexibility in positioning as a
CR plate
ANS OF PREVIOUS BLOG
501.True regarding amorphous silicon thin-film transistor (TFT) arrays
a.TFT is essentially a large integrated
circuit
b.the dimensions of array correspond
to the size of the area to be imaged
c.a transistor is a device that
minimize an electrical signal
d. in TFT, the amplified signal is
stored as an electrical charge
e.The charge can be released by a high
potential and is applied column by column
501.abd------True regarding amorphous silicon thin-film
transistor (TFT) arrays transistor is a device that amplifies an electrical signal and in TFT, the amplified signal is stored as
an electrical charge.The charge can be released by a high potential and is
applied row by row,so that timing of the detected signal determines the
position of the pixel from which it originated. (Chapter 5, page no.86,second
edition (2008),Farr’s Physics for Medical Imaging).