493.True regarding Computed radiography reader.
a.plate is removed from cassette in
reader and scanned by laser
b.optical fibre above the plate is used for scanning
c.rotating mirror above the plate
direct emitted light to photomultiplier tubes
d.single scanning is enough to extract
information
e.exposure of plate
to light done to erase the residual signalANS OF PREVIOUS BLOG
473.True regarding mammography.
a.aims to show microcacifications that
may be 100micrometer or less
b.show tissues having much lower
intrinsic contrast
c.atomic number of glandular tissues
and adipose tissue (approx 7.4 and 6.5 respectively)
d.use of a broad bremsstrahlung
spectrum for imaging
e.target materials used for
mammography are tungsten ,molybdenum(z=42)
and rhodium(z=45)
473.abc---- In mammography, use of a broad bremsstrahlung
spectrum from a tungsten target is far from ideal and now it is standard
practice to make use of the chararcteristic radiation from a lower atomic
energy material giving rise to low energy X- rays (16-22 kev) .Target materials
used for mammography are molybdenum (z=42) and rhodium (z=45),the
former being most common. (Chapter 4,page no.74,second edition (2008),Farr’s
Physics for Medical Imaging).